Patient-Specific Online Dose Verification Based On Transmission Detector Measurements

MEDICAL PHYSICS(2015)

Cited 30|Views5
No score
Abstract
Background and purpose Since IMRT-techniques lead to an increasingly complicated environment, a patient specific IMRT-plan verification is recommended. Furthermore, verifications during patient irradiation and 3D dose reconstruction have the potential to improve treatment delivery, accuracy and safety. This study provides a detailed investigation of the new transmission detector (DTD) Dolphin (IBA Dosimetry, Germany) for online dosimetry. Materials and methods The clinical performance of the DTD was tested by dosimetric plan verification in 2D and 3D for 18 IMRT-sequences. In 2D, DTD measurements were compared to a pre-treatment verification method and a treatment planning system by gamma index and dose difference evaluations. In 3D, dose–volume-histogram (DVH) indices and gamma analysis were evaluated. Furthermore, the error detection ability was tested with leaf position uncertainties and deviations in the linear accelerator (LINAC) output. Results The DTD measurements were in excellent agreement to reference measurements in both 2D (γ3%,3mm=(99.7±0.6)% <1, ΔD±5%=(99.5±0.5)%) and 3D. Only a small dose underestimation (<2%) within the target volume was observed when analyzing DVH-indices. Positional errors of the leaf banks larger than 1mm and errors in LINAC output larger than 2% were identified with the DTD. Conclusions The DTD measures the delivered dose with sufficient accuracy and is therefore suitable for clinical routine.
More
Translated text
Key words
Quality assurance,Transmission detector,Online dose verification
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined