Sci—Thur PM: Imaging — 04: an Iterative Triple Energy Window (TEW) Approach to Cross Talk Correction in Quantitative Small Animal Tc99m and In111 SPECT

Medical physics on CD-ROM/Medical physics(2014)

引用 1|浏览1
暂无评分
摘要
Dual isotope SPECT allows simultaneous measurement of two different tracers in vivo. With In111 (emission energies of 171keV and 245keV) and Tc99m (140keV), quantification of Tc99m is degraded by cross talk from the In111 photons that scatter and are detected at an energy corresponding to Tc99m. TEW uses counts recorded in two narrow windows surrounding the Tc99m primary window to estimate scatter. Iterative TEW corrects for the bias introduced into the TEW estimate resulting from un‐scattered counts detected in the scatter windows. The contamination in the scatter windows is iteratively estimated and subtracted as a fraction of the scatter‐corrected primary window counts. The iterative TEW approach was validated with a small‐animal SPECT/CT camera using a 2.5mL plastic container holding thoroughly mixed Tc99m/In111 activity fractions of 0.15, 0.28, 0.52, 0.99, 2.47 and 6.90. Dose calibrator measurements were the gold standard. Uncorrected for scatter, the Tc99m activity was over‐estimated by as much as 80%. Unmodified TEW underestimated the Tc99m activity by 13%. With iterative TEW corrections applied in projection space, the Tc99m activity was estimated within 5% of truth across all activity fractions above 0.15. This is an improvement over the non‐iterative TEW, which could not sufficiently correct for scatter in the 0.15 and 0.28 phantoms.