9.4-14.1 T small-animal PET-MR imaging: Feasibility analysis of LYSO APD readout via long signal lines

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2013)

引用 0|浏览4
暂无评分
摘要
In the present work we intend to assess the readout feasibility of Avalanche Photodiode (APD) detectors via long signal lines for the development of a combined small-animal PET-MR prototype based on the ClearPEM technology. The detection performance of a LYSO-APD module was evaluated reading out the APD charge signals to the front-end ASIC via an 80mm length flexible flat-cable (FFC). Experimental results showed a time resolution of 5.06ns for the detector module in double-readout mode, with a nonsignificant degradation of 8.4% with the introduction of the FFC. The energy resolution of the system was not degradated by the FFC.
更多
查看译文
关键词
Small-animal PET,ClearPEM,UHF-MRI,Simultaneous PET-MRI