Performance of digital gamma ray spectrometer for loss free counting and its application to NAA using short-lived radionuclides

Journal of Radioanalytical and Nuclear Chemistry(2014)

Cited 1|Views1
No score
Abstract
A digital gamma-ray spectrometer with loss-free counting system was used for real time correction for dead time (DT) due to pulse processing in the electronics. This spectrometer was used for radioactive assay of samples having constant as well as varying DT conditions. The system was tested by measuring activity of short and medium lived nuclides namely 28 Al, 52 V and 128 I in the DT ranges of 80–2 %. Using this spectrometer and neutron activation analysis (NAA), concentrations of Al, V, Ti, Ca, Dy and Mn were determined in some samples and reference materials.
More
Translated text
Key words
Digital gamma ray spectrometer, Short-lived radionuclides, Dead time, Loss free counting, ZDT mode, NAA