EFFECTIVE DOSE MEASUREMENT FOR CONE BEAM COMPUTED TOMOGRAPHY USING GLASS DOSIMETER

Nuclear Engineering and Technology(2014)

引用 18|浏览13
暂无评分
摘要
During image-guided radiation therapy, the patient is exposed to unwanted radiation from imaging devices built into the medical LINAC. In the present study, the effective dose delivered to a patient from a cone beam computed tomography (CBCT) machine was measured. Absorbed doses in specific organs listed in ICRP Publication 103 were measured with glass dosimeters calibrated with kilovolt (kV) X-rays using a whole body physical phantom for typical radiotherapy sites, including the head and neck, chest, and pelvis. The effective dose per scan for the head and neck, chest, and pelvis were 3.37 +/- 0.29, 7.36 +/- 0.33, and 4.09 +/- 0.29 mSv, respectively. The results highlight the importance of the compensation of treatment dose by managing imaging dose.
更多
查看译文
关键词
Effective Dose,Cone Beam Computed Tomography,Image Guided Radiotherapy