Calypso’s array attenuation

JOURNAL OF RADIOTHERAPY IN PRACTICE(2015)

引用 0|浏览11
暂无评分
摘要
Introduction: The Calypso 4D Localization System gives the possibility to track the tumour during treatment, with no additional ionising radiation delivered. To monitor the patient continuously an array is positioned above the patient during the treatment. We intend to study, for various gantry angles, the attenuation effect of the array for 6- and 10 MV and flattening filter free ( FFF) 6- and FFF 10 MV photon beams. Materials and methods: Measurements were performed using an ion chamber placed in a slab phantom positioned at the linac isocenter for 6 MV, 10 MV, FFF 6 MV and FFF 10 MV photon beams. Measurements were performed with and without array above the phantom for 0 degrees, 10 degrees, 20 degrees, 40 degrees and 50 degrees beam angle for a True Beam STx linac, for 5x5 and 10x10 and 15x15 cm(2) field size beams to evaluate the attenuation of the array. A VMAT treatment plan was measured using an ArcCheck with and without the array in the beam path. Results and discussion: Attenuation measured values were up to 3%. Attenuation values were between 1 and 2% with the exception of the 30 degrees- 50 degrees gantry angles which were up to 3.3%. The ratio values calculated in the ArcCheck for relative dose and absolute dose 10 were both 1 . 00. Conclusion: Attenuation of the treatment beam by the Calypso array is within acceptable limits.
更多
查看译文
关键词
ArcCheck,array,attenuation,Calypso