Determination of nodularin using immunoaffinity cartridge-high performance liquid chromatography

ANALYTICAL METHODS(2014)

引用 1|浏览5
暂无评分
摘要
An immuno-antigen of nodularin was prepared, and then a polyclonal antibody against nodularin was obtained from Japanese white rabbits immunized with the immuno-antigen. lmmunoaffinity cartridges were prepared using CNBr-activated Sepharose 4B and the polyclonal antibody against nodularin and their coupling rate ranged from 77.5% to 90.2%. The cartridge load for nodularin and the recovery ranged from 70 ng to 98 ng and 71.8% to 97.2%, respectively. A method for trace level analysis of nodularin in water using the immunoaffinity cartridge and HPLC was established, its detection limit was 3 ng L-1 and the linear range was from 10 to 500 ng L-1.
更多
查看译文
关键词
Neurotoxicity
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要