Evaluation of the Long-Term Stability and Temperature Coefficient of Dew-Point Hygrometers

R. Benyon, T. Vicente,P. Hernández, L. De Rivas, F. Conde

International Journal of Thermophysics(2012)

引用 0|浏览4
暂无评分
摘要
The continuous quest for improved specifications of optical dew-point hygrometers has raised customer expectations on the performance of these devices. In the absence of a long calibration history, users with a limited prior experience in the measurement of humidity, place reliance on manufacturer specifications to estimate long-term stability. While this might be reasonable in the case of measurement of electrical quantities, in humidity it can lead to optimistic estimations of uncertainty. This article reports a study of the long-term stability of some hygrometers and the analysis of their performance as monitored through regular calibration. The results of the investigations provide some typical, realistic uncertainties associated with the long-term stability of instruments used in calibration and testing laboratories. Together, these uncertainties can help in establishing initial contributions in uncertainty budgets, as well as in setting the minimum calibration requirements, based on the evaluation of dominant influence quantities.
更多
查看译文
关键词
Calibration,Dew-point temperature,Drift,Humidity,Hygrometer,Traceability
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要