Diode Laser Cavity Ring-Down Spectroscopy for in Situ Measurement of NO3 Radical in Ambient Air

Journal of Quantitative Spectroscopy & Radiative Transfer/Journal of quantitative spectroscopy & radiative transfer(2015)

引用 22|浏览22
暂无评分
摘要
A cavity ring-down spectroscopy (CRDS) instrument for measuring atmospheric NO3 radical developed in our laboratory is presented in detail. Light from a red laser diode (661.85 um) is coupled on-axis into an optical cavity formed by a pair of high-reflectivity mirrors (R >= 99.9985%) to achieve an effective absorption path length of approximately 20 km. The detection limit of the NO3 radical determined by Allan variance for the field observation with high particles is approximately 3.2 pptv (2 sigma, 10 s). The transmission efficiency of the NO3 radical in the system is calibrated, including the filter loss and surface loss. Moreover, measurable interferences from NO2, O-3 and water vapor are also discussed. Considering the influence of inlet transmission efficiency and other factors, the instrument accuracy for NO3 radical measurement is approximately + 8% (1 sigma).The measurement of NO3 radical was performed at a suburb site in Beijing under the situation of high particles concentration (PM2.5 approximately several tens to 150 mu g/m(3)) from October 26 to November 11, 2014. The NO3 radical concentration during the period is relatively low with the maximum value of 38 pptv. The observation results on October 29, combining NO2, O-3 and NO data, are briefly analyzed. The experimental results demonstrate that this compact CRDS instrument has the potential for NO3 radical measurements in the field with high particles. (C) 2015 The Authors. Published by Elsevier Ltd.
更多
查看译文
关键词
NO3 radical,Cavity ring-down spectroscopy (CRDS),Atmospheric monitoring,Diode laser
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要