Background filtering for accuracy improvement in computed tomography with iterative region-of-interest reconstruction

Proceedings of SPIE(2013)

引用 1|浏览0
暂无评分
摘要
Two methods for preventing the deterioration of the accuracy of iterative region-of-interest (ROT) reconstruction are proposed. Both methods apply filters; the first one applies them to the whole region the outside the region of interest (outside ROT) without distinguishing objects ("method 1", hereafter); and the second one applies them to only the air and patient-table regions while masking other objects outside the ROT ("method 2"). The effectiveness of both methods was evaluated in terms of simulated CT values by using two different phantoms. Method 1 reduced the artifact intensity level by 86% (at most) compared with that obtained with the conventional method. In the case of an object with high attenuation coefficient, method 2 decreases the level more than method 1. In other words, method 2 improves reconstruction accuracy without causing deterioration by the filters. By selecting either method 1 or 2 in accordance with the attenuation coefficient in regions of objects to be imaged, it is possible to reduce the error level compared with the conventional method.
更多
查看译文
关键词
Image Reconstruction,Iterative Reconstruction,Region-of-Interest Reconstruction,Targeted Reconstruction,Background Filtering,Accuracy Improvement,Artifacts Reduction,Computed Tomography