���行平板形電離箱(ルース型およびアドバンスト・マーカス型)の校正定数比の実験的評価

Japanese journal of medical physics : an official journal of Japan Society of Medical Physics(2005)

引用 22|浏览7
暂无评分
摘要
A Japanese code of practice for clinical dosimetry, titled ''Standard Dosimetry of Absorbed Dose in External Beam Radiotherapy'' was published by the Japan Society of Medical Physics (JSMP) in 2002. It mostly followed International Atomic Energy Agency (IAEA) Technical Reports Series No.398, which was based on N{sub D,w}, i.e., the calibration factor in terms of absorbed dose to water for a dosimeter. The Japanese primary standard dosimetry laboratory, however, has not supplied N{sub D,w} but N{sub x}, as the calibration factor in terms of exposure. The unique feature of the Japanese code was provision of a data table of calculated conversion factors, N{sub D,w}/N{sub x} values, for many types of ionization chambers, excluding new plane-parallel ionization chambers. This paper describes the experimental evaluation of the conversion factors for the new plane-parallel ionization chambers, such as the Roos-type and Advanced Markus chambers. The obtained N{sub D,w}/N{sub x} values for PTW 34001, Wellhoefer PPC 40 and PTW 34045 were 37.96{+-}0.19, 37.85{+-}0.36 and 37.90{+-}0.26 (Gy/C kg{sup -1}), respectively. They agreed with estimations based on Monte Carlo calculations. (author)