An FPGA-based lock-in detection system to enable Chemical Species Tomography using TDLAS

2015 IEEE International Conference on Imaging Systems and Techniques (IST)(2015)

引用 8|浏览2
暂无评分
摘要
This paper presents the design, implementation and test of a compact, low-cost and fully digital signal recovery system for tunable diode laser absorption spectroscopy (TDLAS) in narrow line-width gas sensing applications. An FPGA-based digital lock-in amplifier (DLIA), in conjunction with TDLAS using the wavelength modulation spectroscopy (WMS) technique, is utilized to demodulate and extract first (1f) and second (2f) harmonic signals for a narrow CO 2 feature in the spectrum region of 1997.2nm. The spectrum in this wavelength region shows suitably weak water absorption, enabling CO 2 detection with high resolution. Gas-cell experiments were carried out using the DLIA and a conventional rack-mounted commercial lock-in amplifier. The comparison between the two systems shows good agreement, validating the feasibility of this approach and demonstrating the prospect for extension to a massively multichannel system to implement Chemical Species Tomography.
更多
查看译文
关键词
Field Programmable Gate Arrays (FPGA),Lock-in amplifier,Digital demodulation,Tunable Diode Laser Spectroscopy (TDLAS),Chemical Species Tomography (CST)