Experimental calibration of x-ray camera performance: spatial resolution, flat field response, and radiation sensitivity

Proceedings of SPIE(2014)

引用 0|浏览18
暂无评分
摘要
Major parameters of X-rays camera include spatial resolution, flat field response and dynamic range. Such parameters were calibrated on a pulsed X-rays source with about 0.3MeV energy. Fluorophotometric method was used for the measurement of spatial resolutions of the penetrating lights and reflecting lights. Results indicated they were both basically same. And the spatial resolution of the camera was measured with edge method. Corresponding to 10% intensity, the modulator transfer function (MTS) of the resolution was about 51p/mm, while the size of the point spread function (PSF) was about 0.8mm. Due to the system design with both short distance and big filed of view, the flat field non-homogeneity was about 15%. In addition, because of the relatively big gain of the scintillator and MCP image intensifier and the limited detecting efficiency of the X-rays and scintillator, the image intensity of the flat field response demonstrated a big standard deviation of about 1375. Due to the crosstalk throughout the system, the maximal signal-to-noise ratio (SNR) of the X-rays camera was about 10:1. These results could provide important technical specifications for both applications of X-rays camera and data processing of other relevant images.