Study Of Parameters Influencing The Response Of Radfets

2013 14TH EUROPEAN CONFERENCE ON RADIATION AND ITS EFFECTS ON COMPONENTS AND SYSTEMS (RADECS)(2013)

Cited 23|Views25
No score
Abstract
Fraunhofer INT together with CERN calibrated different RADiation-sensitive Field-Effect Transistor (RADFET) radiation dosimeters from two vendors to be operated in CERN's RadMON system with Co-60 gamma radiation to monitor total ionizing dose (TID) at CERN accelerators [1]. The aim of this study was to find out which RADFET type was best suited for the RadMON system and what parameters influence the radiation-induced threshold voltage shift of the pMOS transistors most. Therefore we looked at the sensor response while changing dose-rate, reading conditions, package, lot-to-lot variation during irradiation. In addition CERN previously examined the effects of different particle or radiation types on their response [2]. Compared to former measurements we reached excellent accuracies which guarantees good dose sensitivity during application at CERN. Additionally the findings of this study can also be of interest for those who will use RADFETs in space or other radiation environments as dosimeters.
More
Translated text
Key words
Terms Characterization, Co-60 gamma radiation, RADFET, radiation dosimeters, total ionizing dose