Dual source CT (DSCT) imaging of obese patients: evaluation of CT number accuracy, uniformity, and noise

Proceedings of SPIE(2009)

引用 0|浏览16
暂无评分
摘要
Obese patients present challenges in obtaining sufficient x-ray exposure over reasonable time periods for acceptable CT image quality. To overcome this limitation, the exposure can be divided between two x-ray sources using a dualsource (DS) CT system. However, cross-scatter issues in DS CT may also compromise image quality. We evaluated a DS CT system optimized for imaging obese patients, comparing the CT number accuracy and uniformity to the same images obtained with a single-source (SS) acquisition. The imaging modes were compared using both solid cylindrical PMMA phantoms and a semi-anthropomorphic thorax phantom fitted with extension rings to simulate different size patients. Clinical protocols were used and CTDIvol and kVp were held constant between SS and DS modes. Results demonstrated good agreement in CT number between SS and DS modes in CT number, with the DS mode showing better axial uniformity for the largest phantoms.
更多
查看译文
关键词
image quality
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要