Comparison of Two High-End Ultrasound Systems for Contrast-Enhanced Ultrasound Quantification of Mural Microvascularity in Crohn's Disease.

ULTRASCHALL IN DER MEDIZIN(2016)

引用 25|浏览18
暂无评分
摘要
Purpose: To verify the reproducibility of contrast-enhanced ultrasound (CEUS) quantification results of two different high-end ultrasound systems and to evaluate the clinical utility of the method in patients with Crohn's disease (CD). Materials and Methods: 18 patients with histologically confirmed CD (36.8% women, 63.2 % men; mean age 43.7 +/- 14.1 years) and wall segments thicker than 5 mm were recruited. CEUS quantification and conventional ultrasound investigation were performed under standardized settings using Toshiba Aplio500 and Siemens Acuson S3000 high-end ultrasound systems. CEUS was performed at a low mechanical index of 0.1 after bolus application of 4.8 ml of SonoVue (R) contrast medium. The recorded DICOM clips were quantified using VueBox (R) (version 4.3) calculating 11 quantitative parameters. Subsequently, CEUS quantification and conventional ultrasound results were analyzed. Results: Correlation of quantitative parameters between the Aplio500 and AcusonS3000 systems for peak enhancement (PE), rise time (RT), wash-in-rate (WiR) and quality of fit (QOF) yielded significance levels of p < 0.05 and p < 0.0001 for wash-in-wash-out area under the curve (WiWoAUC). Spearman rank test showed moderate levels of correlation for PE, RT, WiR and QOF (r = 0.5, 0.49, 0.49 and 0.5 respectively), and high correlation for WiWoAUC (r = 0.89) between the two ultrasound systems. Conclusion: Due to multiple uncontrollable affecting factors, the method of CEUS quantification by VueBox in the intestine cannot be recommended for device-independent multicenter studies. Therefore we suggest to use identical ultrasound systems and probes as well as to establish adequate reference ROIS, like a AIF-ROI.
更多
查看译文
关键词
Crohn's disease,CEUS quantification,contrast agents,utrasonography
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要