Quality-controlled dose-reduction of pelvic X-ray examinations in infants with hip dysplasia
European Journal of Radiology(2016)
Abstract
•Pelvic radiographs of infants with hip dysplasia with speed class 400 and 800 were compared.•Radiographs with reduced dose (800) were not inferior to those with standard dose (400).•We recommend speed class 800 as new reference parameter in these examinations.
MoreTranslated text
Key words
Pelvic X-ray examination,Plain radiography,Radiation dose reduction,Quality management,Radiation protection,Developmental dysplasia of the hip