Quality-controlled dose-reduction of pelvic X-ray examinations in infants with hip dysplasia

European Journal of Radiology(2016)

Cited 9|Views19
No score
Abstract
•Pelvic radiographs of infants with hip dysplasia with speed class 400 and 800 were compared.•Radiographs with reduced dose (800) were not inferior to those with standard dose (400).•We recommend speed class 800 as new reference parameter in these examinations.
More
Translated text
Key words
Pelvic X-ray examination,Plain radiography,Radiation dose reduction,Quality management,Radiation protection,Developmental dysplasia of the hip
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined