谷歌Chrome浏览器插件
订阅小程序
在清言上使用

Ex Vivo Model for Percutaneous Vertebroplasty

Key Engineering Materials(2014)

引用 1|浏览27
暂无评分
摘要
The testing of novel biomaterials for percutaneous vertebroplasty depends on suitable animal models. The aim of this study was to develop ex vivo a reproducible and feasible model of percutaneous vertebroplasty, for ulterior application in vivo . A large animal model was used (Merino sheep), due to its translational properties. Vertebroplasty was performed under tactile and fluoroscopic control, through a bilateral modified parapedicular access in lumbar vertebrae (n=12). Care was taken in order to avoid disruption of the vertebral foramen. The average defect volume was 1234±240 mm 3 . This mean volume ensures practical defects to test novel injectable biomaterials. 6 vertebrae were injected with a commercial cement (Cerament®, Bone Support, Sweden). Adequate defect filling was observed in all vertebrae. All vertebrae were assessed by microCT, prior to and post defect creation and after biomaterial injection. All vertebrae were mechanical tested. No mechanical failure was observed under loads higher than the physiological. Ultimately, this model is considered suitable for pre-clinical in vivo studies, mimicking clinical application.
更多
查看译文
关键词
biomaterial
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要