The aim of this study is to evaluate feasibility and accuracy of dental implant placement utilizing a dedicated bone-supported surgical template. Thirty-eight implants (sixteen in maxilla, twenty-two in mandible) were placed in seven fully edentulous jaws (three maxillae, four mandibles) guided by the designed bone-supported surgical template. A voxel-based registration technique was applied to match pre-and post-operative CBCT scans. The mean angular deviation and mean linear deviation at the implant hex and apex were 6.4 +/- 3.7 degrees(0.7 degrees-14.8 degrees), 1.47 +/- 0.64 mm (0.5-2.56 mm) and 1.70 +/- 1.01 mm (0.71-4.39 mm), respectively. The presented bone-supported surgical template showed acceptable accuracy for clinical use. In return for reduced accuracy, clinicians gain accessibility when using this type of surgical template for both the maxilla and the mandible. This is particularly important in patients with reduced mouth opening.