Circuit Impedance Could be a Crucial Factor Influencing Radiofrequency Ablation Efficacy and Safety: A Myocardial Phantom Study of the Problem and its Correction.

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY(2016)

引用 29|浏览20
暂无评分
摘要
RF Lesion Size and Overheating Are Dependent on Circuit Impedance BackgroundCircuit impedance could affect the safety and efficacy of radiofrequency (RF) ablation. AimTo perform irrigated RF ablations with graded impedance to compare (1) lesion dimensions and overheated dimensions in fixed power ablations (2) and in power corrected ablations. MethodsAblations were performed with irrigated Navistar Thermocool catheter and Stockert EP shuttle generator at settings of 40 W power for 60 seconds, in a previously validated myocardial phantom. The impedance of the circuit was set at 60 , 80 , 100 , 120 , 140 , and 160 . The lesion and overheated dimensions were measured at 53 degrees C and 80 degrees C isotherms, respectively. In the second set of ablations, power was corrected according to circuit impedance. ResultsIn total, 70 ablations were performed. The lesion volume was 72.0 4.8% and 44.7 +/- 4.6% higher at 80 and 100 , respectively, compared to that at 120 and it was 15.4 +/- 1.2%, 28.1 +/- 2.0%, and 38.0 +/- 1.8% lower at 140 , 160 , and 180 , respectively. The overheated volume was four times larger when impedance was reduced to 80 from 100 . It was absent at 120 and above. In the power corrected ablations, the lesion volumes were similar to that of 40 W/120 ablations and there was no evidence of overheating. ConclusionThe lesion and overheated dimensions were significantly larger with lower circuit impedance during irrigated RF ablation and the lesion size was smaller in high impedance ablations. Power delivery adjusted to impedance using a simple equation improved the consistency of lesion formation and prevented overheating.
更多
查看译文
关键词
circuit impedance,lesion dimensions,myocardial phantom,radiofrequency ablation,safety
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要