FADING EFFECT OF LiF:Mg,Ti AND LiF:Mg,Cu,P Ext-Rad AND WHOLE-BODY DETECTORS.

RADIATION PROTECTION DOSIMETRY(2016)

Cited 12|Views33
No score
Abstract
Thermoluminescence dosemeters are widely used in individual and environmental monitoring. The aim of this work was to compare the thermal stability of dosemeters of the Ext-Rad and whole-body card types with LiF:Mg,Ti and LiF:Mg,Cu,P detectors stored at different temperatures and periods. The dosemeters were stored at 0 degrees C, room temperature and 40 degrees C for periods that lasted 8, 30, 45, 90 and 120 d. In general, TLD-100H detectors present higher TL signal stability than TLD-100 detectors. The intensity of the signal remained constant for both materials for storage periods at 0 degrees C. At RT the same results was observed for TLD-100H. For TLD-100 detectors, a maximum variation of 22 % was registered for the longest period. At 40 degrees C the TL signal decreased with storage time for both detectors. The TL signal of TLD-100H detectors presented maximum variations of 12 % whereas for TLD-100 detectors, larger variations of 25 % were observed.