Pulse Oximetry: A Non-Invasive, Novel Marker For The Quality Of Chest Compressions In Porcine Models Of Cardiac Arrest

PLOS ONE(2015)

Cited 12|Views34
No score
Abstract
ObjectivePulse oximetry, which noninvasively detects the blood flow of peripheral tissue, has achieved widespread clinical use. We have noticed that the better the quality of cardiopulmonary resuscitation (CPR), the better the appearance of pulse oximetry plethysmographic waveform (POP). We investigated whether the area under the curve (AUC) and/or the amplitude (Amp) of POP could be used to monitor the quality of CPR.DesignProspective, randomized controlled study.SettingAnimal experimental center in Peking Union Medical Collage Hospital, Beijing, China.SubjectsHealthy 3-month-old male domestic swine.Interventions34 local pigs were enrolled in this study. After 4 minutes of untreated ventricular fibrillation, animals were randomly assigned into two resuscitation groups: a "low quality" group (with a compression depth of 3cm) and a "high quality" group (with a depth of 5cm). All treatments between the two groups were identical except for the depth of chest compressions. Hemodynamic parameters [coronary perfusion pressure (CPP), partial pressure of end-tidal carbon dioxide (PETCO2)] as well as AUC and Amp of POP were all collected and analyzed.Measurements and FindingsThere were statistical differences between the "high quality" group and the "low quality" group in AUC, Amp, CPP and PETCO2 during CPR (P<0.05). AUC, Amp and CPP were positively correlated with PETCO2, respectively (P<0.01). There was no statistical difference between the heart rate calculated according to the POP (F-CPR) and the frequency of mechanical CPR at the 3rd minute of CPR. The F-CPR was lower than the frequency of mechanical CPR at the 6th and the 9th minute of CPR.ConclusionsBoth the AUC and Amp of POP correlated well with CPP and PETCO2 in animal models. The frequency of POP closely matched the CPR heart rate. AUC and Amp of POP might be potential noninvasive quality monitoring markers for CPR.
More
Translated text
Key words
blood pressure,compression,signal filtering,hemodynamics
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined