Liquid chromatography-tandem mass spectrometry for the quantification of flurbiprofen in human plasma and its application in a study of bioequivalence.

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences(2015)

Cited 14|Views19
No score
Abstract
A simple, quick and accurate LC-MS/MS method for the quantification of flurbiprofen in human plasma with indomethacin as internal standard (IS) was developed and validated. Samples were treated with methanol to precipitate proteins, then separated on a Ultimate C18 column (5μm, 2.1×50mm) with a gradient elusion process. Mobile phase A was comprised of water and formic acid, mobile phase B was comprised of acetonitrile and formic acid. Multi reaction monitoring (MRM) signals were saved on a negative ionization electrospray mass spectrometer. The calibration curve showed good linearity in the range of 40.00-10000.00μg/L (r(2)=0.998). Intra-day RE was 0.2-2.2%. Inter-day RE was 0.5-3.4%. The samples showed good stability under the study conditions. No significant matrix effect was observed. The established method was then applied to a bioequivalence study of a flurbiprofen axetil formulation.