Implant-Abutment Connection Deformation After Prosthetic Procedures: An In Vitro Study

INTERNATIONAL JOURNAL OF PROSTHODONTICS(2015)

引用 8|浏览3
暂无评分
摘要
This study tested the possible damage to the internal implant connection provoked by repeated disconnection and reconnection of prosthetic components. Using a light-structured scanner, connection deformation was inferred by three-dimensional (3D) positional changes of a "reference" abutment before and after multiple dis- and reconnections. Measurements were taken after 1, 2, 5, 10, and 20 insertions of titanium abutments in 16 internal hexagon implants. Statistical analysis revealed that multiple dis- and reconnections could cause a deformation in the implant-abutment connection, proportional to the number of insertion procedures. However, below the threshold of 10 cycles, 3D deformation values were minimal.
更多
查看译文
关键词
prosthetic procedures,deformation,implant-abutment
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要