Development Of A Smartphone-Based Reading System For Lateral Flow Immunoassay

JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY(2014)

引用 17|浏览1
暂无评分
摘要
This study was conducted to develop and evaluate the performance of the Smartphone-based reading system for the lateral flow immunoassay (LFIA). Smartphone-based reading system consists of a Samsung Galaxy S2 Smartphone, Smartphone application, and a LFIA reader. LFIA reader is composed of the close-up lens with a focal length up to 30 mm, white LED light, lithium polymer battery, and main body. The Smartphone application for image acquisition and data analysis was developed on the Android platform. The standard curve was obtained by plotting the measured P-T/P-C or A(T)/A(C) ratio versus Salmonella standard concentration. The mean, standard deviation (SD), recovery, and relative standard deviation (RSD) were also calculated using additional experimental results. These data were compared with that obtained from the benchtop LFIA reader. The LOD in both systems was observed with 10(6) CFU/mL. The results show high accuracy and good reproducibility with a RSD less than 10% in the range of 10(6) to 10(9) CFU/mL. Due to the simple structure, good sensitivity, and high accuracy of the Smartphone-based reading system, this system can be substituted for the benchtop LFIA reader for point-of-care medical diagnostics.
更多
查看译文
关键词
Smartphone,Lateral Flow Immunoassay,Salmonella
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要