A clinical biomarker assay to quantitate thymic stromal lymphopoietin in human plasma at sub-pg/ml level.

BIOANALYSIS(2015)

引用 2|浏览21
暂无评分
摘要
Background: Thymic stromal lymphopoietin (TSLP) is an attractive therapeutic target for the treatment of allergic diseases, and plasma TSLP is a potential patient selection marker in the development of therapeutic agents. Results: We developed and validated an ultrasensitive electrochemiluminescence assay for measurement of TSLP in plasma with a lower limit of quantitation of 0.12 pg/ml, which allowed the quantitation of TSLP in approximately 90% of human plasma samples tested. The assay demonstrated excellent performance characteristics, including precision, accuracy, sensitivity and dilution linearity. Stability and biological variability of TSLP in plasma were also assessed for clinical sample analysis and data interpretation. Conclusion: The validated TSLP assay enables assessment of circulating TSLP as a patient selection marker in the development of therapeutics to treat atopic diseases.