Maintaining the accuracy of the 60 Co calibration service at the ARPANSA post source replacement in 2010

AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE(2015)

Cited 1|Views4
No score
Abstract
The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) maintains a 60 Co teletherapy source primarily for the calibration of therapy dosemeters. The source and encapsulating head were replaced in early 2010 with an Eldorado 78 head and new 60 Co source. In this article we present the results of ongoing accuracy and stability measurements since the replacement. A number of formal and informal indirect comparisons have been carried out with laboratories holding primary and secondary standards for 60 Co. ARPANSA chambers have also been calibrated at international primary standard laboratories allowing comparison of calibration coefficients and thus 60 Co absorbed dose standards. 60 Co calibration coefficients supplied by manufacturers of chambers were compared to those measured at the ARPANSA when this calibration was traceable to a primary standard. ARPANSA also participates in an annual international mailed dosimetry audit conducted by the International Atomic Energy Agency. The results thus far demonstrate that the absorbed doses to water delivered by the new ARPANSA 60 Co source are consistent with international doses within the stated uncertainties.
More
Translated text
Key words
60Co,Calibration,Comparison,Primary standard,Quality control,Quality assurance