Edge Thomson scattering diagnostic on COMPASS tokamak: installation, calibration, operation, improvements.

REVIEW OF SCIENTIFIC INSTRUMENTS(2014)

Cited 13|Views18
No score
Abstract
The core Thomson scattering diagnostic (TS) on the COMPASS tokamak was put in operation and reported earlier. Implementation of edge TS, with spatial resolution along the laser beam up to similar to 1/100 of the tokamak minor radius, is presented now. The procedure for spatial calibration and alignment of both core and edge systems is described. Several further upgrades of the TS system, like a triggering unit and piezo motor driven vacuum window shutter, are introduced as well. The edge TS system, together with the core TS, is now in routine operation and provides electron temperature and density profiles.