Absolute intensity calibration of the 32-channel heterodyne radiometer on experimental advanced superconducting tokamak.

REVIEW OF SCIENTIFIC INSTRUMENTS(2014)

引用 16|浏览4
暂无评分
摘要
This paper presents the results of the in situ absolute intensity calibration for the 32-channel heterodyne radiometer on the experimental advanced superconducting tokamak. The hot/cold load method is adopted, and the coherent averaging technique is employed to improve the signal to noise ratio. Measured spectra and electron temperature profiles are compared with those from an independent calibrated Michelson interferometer, and there is a relatively good agreement between the results from the two different systems. (C) 2014 AIP Publishing LLC.