Application of GEANT4 simulation on calibration of HPGe detectors for cylindrical environmental samples.

JOURNAL OF RADIOLOGICAL PROTECTION(2014)

引用 10|浏览1
暂无评分
摘要
The determination of radionuclide activity concentration requires a prior knowledge of the full-energy peak (FEP) efficiency at all photon energies for a given measuring geometry. This problem has been partially solved by using procedures based on Monte Carlo simulations, developed in order to complement the experimental calibration procedures used in gamma-ray measurements of environmental samples. The aim of this article is to apply GEANT4 simulation for calibration of two HPGe detectors, for measurement of liquid and soil-like samples in cylindrical geometry. The efficiencies obtained using a simulation were compared with experimental results, and applied to a realistic measurement. Measurement uncertainties for both simulation and experimental values were estimated in order to see whether the results of the realistic measurement fall within acceptable limits. The trueness of the result was checked using the known activity of the measured samples provided by IAEA.
更多
查看译文
关键词
HPGe,efficiency,GEANT4