The Motional Stark Effect Polarimeter In The Hl-2a Tokamak

REVIEW OF SCIENTIFIC INSTRUMENTS(2014)

引用 8|浏览11
暂无评分
摘要
A 7-channel motional Stark effect polarimeter based on four polarizers and a spectrometer has been developed in the HL-2A tokamak, which is the first time successful utilizing this kind of polarimeter on a tokamak. The accuracy of the angle can reach +/- 0.25 degrees in the calibration experiments. Pilot experiments of measuring the magnetic pitch angle have been successfully carried out in the weak motional Stark effect plasma discharge with toroidal magnetic field of similar to 1.3 T and beam energy of similar to 25 keV/amu. The pitch angles of magnetic field are obtained for 7 spatial points covering 24 cm along major radius with time resolution of 40 ms; the profiles of safety factor are obtained by combining with the Equilibrium and Reconstruction Fitting Code. The core value of safety factor (q) is less than 1 during the sawtooth oscillation and the position of q = 1 surface is well consistent with the results measured by soft X-ray array. (C) 2014 AIP Publishing LLC.