Calibration Of A Gated Flat Field Spectrometer As A Function Of X-Ray Intensity

REVIEW OF SCIENTIFIC INSTRUMENTS(2014)

引用 8|浏览53
暂无评分
摘要
We present an experimental determination of the response of a gated flat-field spectrometer at the Shenguang-II laser facility. X-rays were emitted from a target that was heated by laser beams and then were divided into different intensities with a step aluminum filter and collected by a spectrometer. The transmission of the filter was calibrated using the Beijing Synchrotron Radiation Facility. The response characteristics of the spectrometer were determined by comparing the counts recorded by the spectrometer with the relative intensities of the x-rays transmitted through the step aluminum filter. The response characteristics were used to correct the transmission from two shots of an opacity experiment using the same samples. The transmissions from the two shots are consistent with corrections, but discrepant without corrections. (C) 2014 AIP Publishing LLC.