Calibration of a 32 channel electron cyclotron emission radiometer on the HL-2A tokamak.

REVIEW OF SCIENTIFIC INSTRUMENTS(2014)

引用 64|浏览8
暂无评分
摘要
A novel 32-channel electron cyclotron emission radiometer has been designed and tested for the measurement of electron temperature profiles on the HL-2A tokamak. This system is based on the intermediate frequency filter detection technique, and has the features of wide working frequency range and high spatial resolution. Two relative calibration methods have been investigated: sweeping the toroidal magnetic field and hopping the output frequency of the local oscillator. Preliminary results show that both methods can ensure reasonable profiles. (C) 2014 AIP Publishing LLC.