Improved spatial resolution and lower-dose pediatric CT imaging: a feasibility study to evaluate narrowing the X-ray photon energy spectrum

Mark G. Benz, Matthew W. Benz, Steven B. Birnbaum,Eric Chason,Brian W. Sheldon,Dale McGuire

Pediatric radiology(2014)

引用 0|浏览1
暂无评分
摘要
This feasibility study has shown that improved spatial resolution and reduced radiation dose can be achieved in pediatric CT by narrowing the X-ray photon energy spectrum. This is done by placing a hafnium filter between the X-ray generator and a pediatric abdominal phantom. A CT system manufactured in 1999 that was in the process of being remanufactured was used as the platform for this study. This system had the advantage of easy access to the X-ray generator for modifications to change the X-ray photon energy spectrum; it also had the disadvantage of not employing the latest post-imaging noise reduction iterative reconstruction technology. Because we observed improvements after changing the X-ray photon energy spectrum, we recommend a future study combining this change with an optimized iterative reconstruction noise reduction technique.
更多
查看译文
关键词
CT imaging, Pediatric, Spatial resolution, Dose reduction, Hafnium filtration, Hf filter, Reduced radiation dose