Determination of the gamma emission intensities of ¹¹¹Ag.

Applied Radiation and Isotopes(2014)

引用 7|浏览2
暂无评分
摘要
A radioactive solution of 111Ag, standardised by the absolute measurement methods 4π(PC)-γ and 4π(LS)-γ coincidence counting at the National Physical Laboratory (NPL), was measured by two independently calibrated HPGe γ spectrometers in order to estimate the γ emission intensities and to determine the absolute intensity, with the aim of improving the currently published values. An absolute intensity value of 6.68 (7)% was obtained for the 342.1keV γ emission, which is in agreement with previously reported values, but greatly reduces the uncertainty. Additionally, this work proposes a new emission intensity for the 450.0keV γ emission that has not been previously reported, with an absolute intensity of 0.000482 (12)%. An investigation of the published γ emission intensities shows significant discrepancies that require resolution.
更多
查看译文
关键词
Ag-111,Gamma emission intensities,Spectroscopy,Radioactivity