[Evaluation of portable TGDc-01 tonometers and comparison with the Goldmann applanation tonometer].

F Rombold,M J Thiel, A S Neubauer, C Hirneiss, A Kampik

Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft(2005)

引用 9|浏览1
暂无评分
摘要
PURPOSE:The aim of this study was to evaluate patients' acceptance and intraocular pressure (IOP) readings of a new digital mobile tonometer (TGDc-01) and compare it to Goldmann applanation tonometry. METHOD:Measurements repeated five times with the TGDc-01 and three times with Goldmann tonometry were performed in 100 eyes of 100 patients by two independent investigators. Patients' acceptance of both techniques was evaluated by a visual analogue scale (VAS). RESULTS:The mean IOP with the TGDc-01 yielded 15.4 mmHg for investigator 1 and 12.7 mmHg for investigator 2 (range: 4-43 mmHg). Results of the measurements with Goldmann tonometry showed 17.6 mmHg for investigator 1 and 17.3 mmHg for investigator 2 (9-42 mmHg). The IOP difference of the two tonometry methods was highly significant (p<0.001). The intraobserver variability was 29% for investigator 1 and 8% for investigator 2. Mean IOP values of the two investigators taken with the TGDc-01 differed significantly (p<0.01) from each other by a mean of 2.6 mmHg. CONCLUSIONS:The new mobile tonometer TGDc-01 is better accepted by patients but IOP values are significantly lower compared to Goldmann tonometry and variability is high. Regarding glaucoma diagnostics it seems to be less suitable than Goldmann tonometry.