Knotless single-row rotator cuff repair: a comparative biomechanical study of 2 knotless suture anchors.

ORTHOPEDICS(2013)

引用 6|浏览1
暂无评分
摘要
The purpose of this study was to compare the gap formation during cyclic loading, maximum repair strength, and failure mode of single-row full-thickness supraspinatus repairs performed using 2 knotless suture anchors with differing internal suture-retention mechanisms in a human cadaver model. Nine matched pairs of cadaver shoulders were used. Full-thickness tears were induced by detaching the supraspinatus tendon from the greater tuberosity. Single-row repairs were performed with either type I (Opus Magnum PI; ArthroCare, Austin, Texas) or type II (ReelX STT; Stryker, Mahwah, New Jersey) knotless suture anchors. The repaired tendon was cycled from 10 to 90 N for 500 cycles, followed by load to failure. Gap formation was measured at 5, 100, 200, 300, 400, and 500 cycles with a video digitizing system. Anchor type or location (anterior or posterior) had no effect on gap formation during cyclic loading regardless of position (anterior, P=.385; posterior, P=.389). Maximum load to failure was significantly greater (P=.018) for repairs performed with type II anchors (288 +/- 62 N) compared with type I anchors (179 +/- 39 N). Primary failure modes were anchor pullout and tendon tearing for type II anchors and suture slippage through the anchor for type I anchors. The internal ratcheting suture-retention mechanism of type II anchors may have helped this anchor outperform the suture-cinching mechanism of type I anchors by supporting significantly higher loads before failure and minimizing suture slippage, potentially leading to stronger repairs clinically.
更多
查看译文
关键词
comparative biomechanical study,single-row
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要