Contaminated Ventilator Air Flow Sensor Linked To Bacillus Cereus Colonization Of Newborns

EMERGING INFECTIOUS DISEASES(2013)

Cited 13|Views3
No score
Abstract
We investigated Bacillus cereus positive tracheal aspirates from infants on ventilators in a neonatal intensive care unit. Multi locus sequence typing determined a genetic match between strains isolated from samples from a case-patient and from the air flow sensor in the ventilator. Changing the sterilization method for sensors to steam autoclaving stopped transmission.
More
Translated text
Key words
Bacillus cereus,air flow sensor,bacteria,cluster,colonization,disinfection,infant,neonate,newborn,ventilator