Validation of the in vivo volumetric wear measurement for total knee prostheses in model-based RSA.

Journal of Biomechanics(2013)

引用 3|浏览3
暂无评分
摘要
Implant failure related to polyethylene wear remains an important issue in total knee arthroplasty. Polyethylene wear is usually assessed in vivo by measuring the remaining insert thickness on X-ray images of the knee. To reflect the amount of wear debris more accurately, a 3-dimensional overlap measurement has been suggested, which is based on implant component models which are matched on calibrated stereo X-ray images using model-based roentgen stereophotogrammatic analysis. The goal of this study was to determine the influence of pose estimation, insert thickness deviation and variation in the femoral-tibial contact location on the accuracy and precision of the measurement using simulations and a phantom experiment.
更多
查看译文
关键词
Total knee arthroplasty,Polyethylene wear,Image-based measurements,Model-based RSA