A free-field method to calibrate bone conduction transducers.

JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA(2013)

引用 9|浏览2
暂无评分
摘要
Bone conduction communication systems employ a variety of transducers with different physical and electroacoustic properties, and these transducers may be worn at various skull locations. Testing these systems thus requires a reliable means of transducer calibration that can be implemented across different devices, skull locations, and settings. Unfortunately, existing calibration standards do not meet these criteria. Audiometric bone conduction standards focus on only one device model and on limited skull locations. Furthermore, while mechanical couplers may be used for calibration, the general human validity of their results is suspect. To address the need for more flexible, human-centered calibration methods, the authors investigated a procedure for bone transducer calibration, analogous to free-field methods for calibrating air conduction headphones. Participants listened to1s third-octave noise bands (125-12 500 Hz) alternating between a bone transducer and a loudspeaker and adjusted the bone transducer to match the perceived loudness of the loudspeaker at each test frequency. Participants tested two transducer models and two skull locations. Intra-and inter-subject reliability was high, and the resulting data differed by transducer, by location, and from the mechanical coupler. The described procedure is flexible to transducer model and skull location, requires only basic equipment, and directly yields perceptual data. [http://dx.doi.org/10.1121/1.4774273]