Commutability Of The Crm 470 C-Reactive Protein Value In The Dade Behring N High Sensitivity Crp Assay

CLINICAL CHEMISTRY AND LABORATORY MEDICINE(2003)

引用 9|浏览3
暂无评分
摘要
Certified Reference Material 470 (CRM 470) demonstrates commutability with both the manufacturers calibrator and with dilutions of serum pools in the Dade Behring N High Sensitivity assay for Creactive protein (CRP). Both regression and back calibration show similar nonlinearity for all materials, largely due to the method of calibration curve fitting used in this assay. Significant differences in values among the currently available commercial assays can be largely overcome by using appropriate calibration curve fitting and the recommended value transfer protocol, which includes a minimum of two assay runs on each of at least 3 separate days, with weight correction of all reconstitutions and dilutions. An initial weightcorrected dilution should be made each day because of the relatively high level of CRP in CRM 470. In our opinion, the degree of nonlinearity, imprecision, and differences in values in currently available assays renders the use of fixed clinical decision cutpoints questionable for highsensitivity CRP. An alternative approach is suggested.
更多
查看译文
关键词
C-reactive protein, reference material, standardization