20 MHz forward-imaging single-element beam steering with an internal rotating variable-angle reflecting surface: Wire phantom and ex vivo pilot study.

Ultrasonics(2013)

Cited 5|Views1
No score
Abstract
Feasibility is demonstrated for a forward-imaging beam steering system involving a single-element 20 MHz angled-face acoustic transducer combined with an internal rotating variable-angle reflecting surface (VARS). Rotation of the VARS structure, for a fixed position of the transducer, generates a 2-D angular sector scan. If these VARS revolutions were to be accompanied by successive rotations of the single-element transducer, 3-D imaging would be achieved. In the design of this device, a single-element 20 MHz PMN-PT press-focused angled-face transducer is focused on the circle of midpoints of a micro-machined VARS within the distal end of an endoscope. The 2-D imaging system was tested in water bath experiments with phantom wire structures at a depth of 10 mm, and exhibited an axial resolution of 66 mu m and a lateral resolution of 520 mu m. Chirp coded excitation was used to enhance the signal-to-noise ratio, and to increase the depth of penetration. Images of an ex vivo cow eye were obtained. This VARS-based approach offers a novel forward-looking beam-steering method, which could be useful in intra-cavity imaging. (C) 2012 Elsevier B.V. All rights reserved.
More
Translated text
Key words
Medical beam and medical steering,Medical ultrasonics,Chirp coded excitation,Acoustic imaging needle,Variable angle reflecting surface