Influences of microbubble diameter and ultrasonic parameters on in vitro sonothrombolysis efficacy.

Journal of Vascular and Interventional Radiology(2012)

引用 28|浏览1
暂无评分
摘要
To quantify the effects of microbubble (MB) size, elasticity, and pulsed ultrasonic parameters on in vitro sonothrombolysis (ultrasound [US]-mediated thrombolysis) efficacy.Monodispersive MBs with diameters of 1 μm or 3 μm were exposed to pulsed US (1 MHz or 3 MHz) to lyse rabbit blood clots. Sonothrombolysis efficacy (clot mass loss) was measured as functions of MB size and concentration, ultrasonic frequency and intensity, pulse duration (PD), pulse repeat frequency (PRF), and duty factor.Sonothrombolysis at 1 MHz was more effective using 3-μm MBs and at 3 MHz using 1-μm MBs. Sonothrombolysis was more effective at 1 MHz when≥75% of MBs remained intact, especially for 3-μm MBs; improving sonothrombolysis by increasing PRF from 100 Hz to 400 Hz at 3 MHz was associated with increasing 3-μm MB survival. However, 60% of 1-μm MBs were destroyed during maximal sonothrombolysis at 3 MHz, indicating that considerable MB collapse may be required for sonothrombolysis under these conditions.The ability to control MB size and elasticity permits using a wide range of US parameters (eg, frequency, intensity) to produce desired levels of sonothrombolysis. Comparable, maximal sonothrombolysis efficacy was achieved at 20-fold lower intensity with 3-μm MBs (0.1W/cm(2)) than with 1-μm MBs (2.0W/cm(2)), a potential safety issue for in vivo sonothrombolysis. US parameters that maximized MB survival yielded maximal sonothrombolysis efficacy except with 1-μm MBs at 3MHz where most MBs were destroyed.
更多
查看译文
关键词
AFM,MB,PBS,PD,PRF,tPA
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要