Three-dimensional brain phantom containing bone and grey matter structures with a realistic head contour

Annals of nuclear medicine(2012)

引用 70|浏览9
暂无评分
摘要
Introduction A physical 3-dimensional phantom that simulates PET/SPECT images of static regional cerebral blood flow in grey matter with a realistic head contour has been developed. This study examined the feasibility of using this phantom for evaluating PET/SPECT images. Methods The phantom was constructed using a transparent, hydrophobic photo-curable polymer with a laser-modelling technique. The phantom was designed to contain the grey matter, the skull, and the trachea spaces filled with a radioactive solution, a bone-equivalent solution of K 2 HPO 4 , and air, respectively. The grey matter and bone compartments were designed to establish the connectivity. A series of experiments was performed to confirm the accuracy and reproducibility of the phantom using X-ray CT, SPECT, and PET. Results The total weight was 1997 ± 2 g excluding the inner liquid, and volumes were 563 ± 1 and 306 ± 2 mL, corresponding to the grey matter and bone compartments, respectively. The apparent attenuation coefficient averaged over the whole brain was 0.168 ± 0.006 cm −1 for Tc-99 m, which was consistent with the previously reported value for humans (0.168 ± 0.010 cm −1 ). Air bubbles were well removed from both grey-matter and bone compartments, as confirmed by X-ray CT. The phantom was well adapted to experiments using PET and SPECT devices. Conclusion The 3-dimensional brain phantom constructed in this study may be of use for evaluating the adequacy of SPECT/PET reconstruction software programs.
更多
查看译文
关键词
attenuation correction,head,pet
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要