In vivo evaluation of a biolimus eluting nickel titanium self expanding stent with overlapping balloon expandable drug eluting and bare metal stents in a porcine coronary model.

EUROINTERVENTION(2009)

Cited 5|Views3
No score
Abstract
Aims: Long lesions and complex vessel anatomy frequently require the use of overlapping stents to treat a lesion. The purpose of this study was to evaluate the long-term effects of overlapping the Axxess Biolimus A9 (TM) eluting stent (BES) with two of the most commonly used, commercially available drug eluting stents. These stents were compared to BxVelocity bare metal (BMS) stents in a porcine coronary stent-injury model. Methods and results: Nineteen juvenile farm swine, 25-35 kg in weight, 3-6 months in age were utilised. Each animal received an Axxess (TM) stent to their coronary artery as permitted by the individual animal's anatomy. A second stent, either a Cypher (TM), sirolimus eluting stent (SES) or, a Taxus (TM), paclitaxel eluting stent (PES), or a BxVelocity (TM) bare metal stent (BMS) were implanted in an overlapped fashion. The animals were then followed for either 28 or 180 days as specified by a randomisation scheme. At the end of each follow-up period, they were euthenised, and the vessels containing the overlapping stents were harvested, processed into histological sections, and analysed. Compared to bare metal stents, overlapped segments using DES exhibited delayed vascular healing compared to both the proximal and distal non-overlap sites at each of the follow-up time point. Overall, in the non-overlap stent segments, SES induced significantly more inflammation and neointimal hyperplasia compared to PES and BMS. Conclusions: In this study of BMS and two different types of DES overlapped with the Axxess Biolimus A9 (TM) eluting stent, we found that while there was a delay in the degree of vascular healing with DES compared to BMS, the specific type of DES that was overlapped with BES did not affect the behaviour of the overlap zone in terms of most of the histomorphometric measures at 28 or 180 days. This was true whether the stent was drug eluting or bare metal. More inflammation with delayed healing was seen in the SES compared to PES and BMS.
More
Translated text
Key words
Drug eluting stent,Overlap stenting,neointimal hyperplasia,porcine oronary model
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined