Performances of the Vitek MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for rapid identification of bacteria in routine clinical microbiology.

JOURNAL OF CLINICAL MICROBIOLOGY(2012)

引用 115|浏览24
暂无评分
摘要
Rapid and cost-effective matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS)-based systems will replace conventional phenotypic methods for routine identification of bacteria. We report here the first evaluation of the new MALDI-TOF MS-based Vitek MS system in a large clinical microbiology laboratory. This system uses an original spectrum classifier algorithm and a specific database designed for the identification of clinically relevant species. We have tested 767 routine clinical isolates representative of 50 genera and 124 species. Vitek MS-based identifications were performed by means of a single deposit on a MALDI disposable target without any prior extraction step and compared with reference identifications obtained mainly with the VITEK2 phenotypic system; if the identifications were discordant, molecular techniques provided reference identifications. The Vitek MS system provided 96.2% correct identifications to the species level (86.7%), to the genus level (8.2%), or within a range of species belonging to different genera (1.3%). Conversely, 1.3% of isolates were misidentified and 2.5% were unidentified, partly because the species was not included in the database; a second deposit provided a successful identification for 0.8% of isolates unidentified with the first deposit. The Vitek MS system is a simple, convenient, and accurate method for routine bacterial identification with a single deposit, considering the high bacterial diversity studied and as evidenced by the low prevalence of species without correct identification. In addition to a second deposit in uncommon cases, expanding the spectral database is expected to further enhance performances.
更多
查看译文
关键词
bacteria
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要