Transradial coronary brachytherapy with the Novoste Beta-Rail system.

Olivier F. Bertrand,Robert De Larochelli�re,Onil Gleeton, Sylvain Plante, Michel Tessier, Jean Guimond

CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS(2002)

引用 6|浏览12
暂无评分
摘要
We report our initial experience in 10 consecutive patients who underwent transradial coronary brachytherapy for in-stent restenosis using a Sr-90/Y source and the Novoste Beta-Rail system. In all patients, procedures were successfully completed using a right transradial approach. We performed the procedures with the Beta-Rail catheter using 7 Fir (Zuma II, Medtronic, MN, n = 5) or 8 Fr (Cordis, Miami, FL; n = 5) guiding catheters. All lesions were successfully dilated and no additional stent was inserted. We used a 40 mm source (n = 3) or a 60 mm source (n = 7) with manual stepping in four cases. In three cases, we did one stepping, and in one case, we did three steppings, The mean dwell time was 195 +/- 44 sec. The mean delivered dose was 23 +/- 3 Gy at 2 mm distance from the source. No radiation treatment was interrupted. Mean fluoroscopy time was 26 +/- 13 min. Procedural success was achieved in all patients. Three patients had mild CK elevations (< 3 times upper normal limit). All patients were pretreated with clopidogrel (300 mg) and combined treatment with aspirin + clopidogrel is to be continued for at least 1 year. Clinical follow-up up to 3 months has not yielded any complication and all patients have remained free from angina. (C) 2002 Wiley-Liss, Inc.
更多
查看译文
关键词
PTCA,IVUS,radiation,restenosis,isotope
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要