Applicability of a System for fully automated nucleic acid extraction from formalin-fixed paraffin-embedded sections for routine KRAS mutation testing.

DIAGNOSTIC MOLECULAR PATHOLOGY(2012)

引用 6|浏览15
暂无评分
摘要
Due to the approval of various new targeted therapies for the treatment of cancer, molecular pathology laboratories with a diagnostic focus have to meet new challenges: simultaneous handling of a large number of samples, small amounts of input material, and fragmentation of nucleic acids because of formalin fixation. As a consequence, fully automated systems for a fast and standardized extraction of high-quality DNA from formalin-fixed paraffin-embedded (FFPE) tissues are urgently needed. In this study, we tested the performance of a fully automated, high-throughput method for the extraction of nucleic acids from FFPE tissues. We investigated the extraction performance in sections of 5 different tissue types often analyzed in routine pathology laboratories (cervix, colon, liver, lymph node, and lung; n = 340). Furthermore, we compared the quality, labor input, and applicability of the method for diagnostic purposes with those of a laboratory-validated manual method in a clinical setting by screening a set of 45 colorectal adenocarcinoma for the KRAS mutation. Automated extraction of both DNA and RNA was successful in 339 of 340 FFPE samples representing 5 different tissue types. In comparison with a conventional manual extraction protocol, the method showed an overall agreement of 97.7% (95% confidence interval, 88.2%-99.9%) for the subsequent mutational analysis of the KRAS gene in colorectal cancer samples. The fully automated system is a promising tool for a simple, robust, and rapid extraction of DNA and RNA from formalin-fixed tissue. It ensures a standardization of sample processing and can be applied to clinical FFPE samples in routine pathology.
更多
查看译文
关键词
automated DNA extraction,FFPE,KRAS,routine diagnostics,molecular pathology
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要