Point-of-care testing (POCT) of coagulation parameters can help optimize transfusion practice in cardiac surgery. Antifibrinolytic agents may, interfere with the laboratory, and/or POCT coagulation assays This randomized controlled study, compared coagulation parameters obtained from a whole blood POCT coagulation device with a typical laboratory, instrument in cardiac surgery, patients receiving aprotinin, epsilon-aminocaproic acid, or normal saline before undergoing cardiopulmonary, bypass. Aliquots Of arterial blood samples from 42 patients were collected perioperatively, and their prothrombin times (PTs) and activated partial thromboplastin times (aPTTs) were measured by, POCT and laboratory, instrumentation. Linear regression and error analyses were used for the method comparison. For PT the POCT device compared favorably, with the laboratory method. For aPTT, the POCT device did not compare well with the laboratory method. Treatment with antifibrinolytic agents does not interfere with determination of PT.