Chrome Extension
WeChat Mini Program
Use on ChatGLM

Modified map-seeking circuit: use of computer-aided detection in locating postoperative retained foreign bodies.

Journal of Surgical Research(2012)

Cited 2|Views10
No score
Abstract
More than 98% of intra-operative X-rays taken to search for postoperative retained foreign bodies (RFBs) have negative findings; in over 30% of cases of such X-rays, the finding is a false negative. Newer technologies created to find RFBs must not only reduce the false-negative rate, but also must not increase the burden of detecting RFBs. We have introduced the use of computer-aided detection (CAD) to facilitate the detection of RFBs on X-rays utilizing a modified version of map-seeking circuit (MSC) algorithm the referenced map-seeking circuit (RMSC), for our proof-of-concept study for detection of needles in plain abdominal X-rays.Images were obtained by using a portable cassette-based X-ray machine and a C-arm (digital) machine, both of which are commonly used in the operating room. The images obtained using these machines were divided into subimages of approximately 250 × 250 pixels each, for a total of 455 subimages from the cassette-based machine (A) and 365 from the digital machine (B) for use as test samples. Images obtained from A and B were analyzed separately using our modified MSC algorithm with a minimum (τ = 0) and a maximum threshold (τ = 0.5).The automated detection rate (positive predictive value) was 86%, with a false positive/negative rate of 10% to 15% when τ was zero.The CAD-based RMSC algorithm has the potential to improve the accuracy with which RFBs can be found in X-rays. Further research is needed to optimize the detection rate and to identify a wider range of RFBs.
More
Translated text
Key words
computer aided detection,map seeking circuit,machine learning,retained foreign bodies
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined